Technical data

Obsolete and Discontinued Commands 7
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide 691
:MEASure:VSTArt
(see page 750)
Command Syntax
:MEASure:VSTArt <vstart_argument>
<vstart_argument> ::= value for vertical marker 1
The :MEASure:VSTArt command moves the vertical marker (Y1 cursor) to
the specified value corresponding to the selected source. The source can
be selected by the MARKer:X1Y1source command.
Query Syntax
:MEASure:VSTArt?
The :MEASure:VSTArt? query returns the current value of the Y1 cursor.
Return Format
<value><NL>
<value> ::= voltage at voltage marker 1 in NR3 format
See Also "Introduction to :MARKer Commands" on page 275
"Introduction to :MEASure Commands" on page 290
":MARKer:Y1Position" on page 282
":MARKer:Y2Position" on page 283
":MARKer:YDELta" on page 284
":MARKer:X1Y1source" on page 278
":MEASure:SOURce" on page 315
":MEASure:TDELta" on page 681
":MEASure:TSTArt" on page 685
NOTE
The short form of this command, VSTA, does not follow the defined Long Form to Short
Form Truncation Rules (see page 752). The normal short form, VST, would be the same for
both VSTArt and VSTOp, so sending VST for the VSTArt command produces an error.
NOTE
The :MEASure:VSTArt command is an obsolete command provided for compatibility to
previous oscilloscopes. Use the :MARKer:Y1Position command (see page 282) instead.