Technical data

Obsolete and Discontinued Commands 7
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide 685
:MEASure:TSTArt
(see page 750)
Command Syntax
:MEASure:TSTArt <value> [suffix]
<value> ::= time at the start marker in seconds
[suffix] ::= {s | ms | us | ns | ps}
The :MEASure:TSTArt command moves the start marker (X1 cursor) to the
specified time with respect to the trigger time.
Query Syntax
:MEASure:TSTArt?
The :MEASure:TSTArt? query returns the time at the start marker (X1
cursor).
Return Format
<value><NL>
<value> ::= time at the start marker in NR3 format
See Also "Introduction to :MARKer Commands" on page 275
"Introduction to :MEASure Commands" on page 290
":MARKer:X1Position" on page 277
":MARKer:X2Position" on page 279
":MARKer:XDELta" on page 281
":MEASure:TDELta" on page 681
":MEASure:TSTOp" on page 686
NOTE
The short form of this command, TSTA, does not follow the defined Long Form to Short
Form Truncation Rules (see page 752). The normal short form "TST" would be the same for
both TSTArt and TSTOp, so sending TST for the TSTArt command produces an error.
NOTE
The :MEASure:TSTArt command is an obsolete command provided for compatibility to
previous oscilloscopes. Use the :MARKer:X1Position command (see page 277) instead.