Technical data

Commands by Subsystem 5
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide 605
:ACQuire:TYPE must be set to NORMal, AVERage, or HRESolution. If
AVERage, :ACQuire:COUNt must be set to 1 in order to return more
than the measurement record or precision analysis record.
MAXimum or RAW will allow up to 8,000,000 points to be returned. The
number of points returned will vary as the instrument's configuration is
changed. Use the :WAVeform:POINts? MAXimum query to determine the
maximum number of points that can be retrieved at the current
settings.
Query Syntax
:WAVeform:POINts:MODE?
The :WAVeform:POINts:MODE? query returns the current points mode.
Setting the points mode will affect what data is transferred. See the
discussion above.
Return Format
<points_mode><NL>
<points_mode> ::= {NORMal | MAXimum | RAW}
See Also "Introduction to :WAVeform Commands" on page 592
":WAVeform:DATA" on page 599
":ACQuire:POINts" on page 184
":SYSTem:PRECision" on page 424
":WAVeform:VIEW" on page 618
":WAVeform:PREamble" on page 606
":WAVeform:POINts" on page 602
":TIMebase:MODE" on page 431
":ACQuire:TYPE" on page 191
":ACQuire:COUNt" on page 181