Technical data
Commands by Subsystem 5
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide 605
• :ACQuire:TYPE must be set to NORMal, AVERage, or HRESolution. If
AVERage, :ACQuire:COUNt must be set to 1 in order to return more
than the measurement record or precision analysis record.
• MAXimum or RAW will allow up to 8,000,000 points to be returned. The
number of points returned will vary as the instrument's configuration is
changed. Use the :WAVeform:POINts? MAXimum query to determine the
maximum number of points that can be retrieved at the current
settings.
Query Syntax
:WAVeform:POINts:MODE?
The :WAVeform:POINts:MODE? query returns the current points mode.
Setting the points mode will affect what data is transferred. See the
discussion above.
Return Format
<points_mode><NL>
<points_mode> ::= {NORMal | MAXimum | RAW}
See Also • "Introduction to :WAVeform Commands" on page 592
• ":WAVeform:DATA" on page 599
• ":ACQuire:POINts" on page 184
• ":SYSTem:PRECision" on page 424
• ":WAVeform:VIEW" on page 618
• ":WAVeform:PREamble" on page 606
• ":WAVeform:POINts" on page 602
• ":TIMebase:MODE" on page 431
• ":ACQuire:TYPE" on page 191
• ":ACQuire:COUNt" on page 181