Technical data

Commands by Subsystem 5
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide 601
:WAVeform:FORMat
(see page 750)
Command Syntax
:WAVeform:FORMat <value>
<value> ::= {WORD | BYTE | ASCii}
The :WAVeform:FORMat command sets the data transmission mode for
waveform data points. This command controls how the data is formatted
when sent from the oscilloscope.
ASCii formatted data converts the internal integer data values to real
Y- axis values. Values are transferred as ASCii digits in floating point
notation, separated by commas.
ASCII formatted data is transferred ASCii text.
WORD formatted data transfers 16- bit data as two bytes. The
:WAVeform:BYTeorder command can be used to specify whether the
upper or lower byte is transmitted first. The default (no command sent)
is that the upper byte transmitted first.
BYTE formatted data is transferred as 8- bit bytes.
Query Syntax
:WAVeform:FORMat?
The :WAVeform:FORMat query returns the current output format for the
transfer of waveform data.
Return Format
<value><NL>
<value> ::= {WORD | BYTE | ASC}
See Also "Introduction to :WAVeform Commands" on page 592
":WAVeform:BYTeorder" on page 597
":WAVeform:DATA" on page 599
":WAVeform:PREamble" on page 606
Example Code "Example Code" on page 611