Technical data

570 Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
:TRIGger:UART Commands
Table 84 :TRIGger:UART Commands Summary
Command Query Options and Query Returns
:TRIGger:UART:BASE
<base> (see page 572)
:TRIGger:UART:BASE?
(see page 572)
<base> ::= {ASCii | HEX}
:TRIGger:UART:BAUDrat
e <baudrate> (see
page 573)
:TRIGger:UART:BAUDrat
e? (see page 573)
<baudrate> ::= integer from 1200
to 3000000 in 100 b/s increments
:TRIGger:UART:BITorde
r <bitorder> (see
page 574)
:TRIGger:UART:BITorde
r? (see page 574)
<bitorder> ::= {LSBFirst |
MSBFirst}
:TRIGger:UART:BURSt
<value> (see
page 575)
:TRIGger:UART:BURSt?
(see page 575)
<value> ::= {OFF | 1 to 4096 in
NR1 format}
:TRIGger:UART:DATA
<value> (see
page 576)
:TRIGger:UART:DATA?
(see page 576)
<value> ::= 8-bit integer from
0-255 (0x00-0xff) in decimal,
<hexadecimal>, <binary>, or
<quoted_string> format
<hexadecimal> ::= #Hnn where n
::= {0,..,9 | A,..,F} for
hexadecimal
<binary> ::= #Bnn...n where n ::=
{0 | 1} for binary
<quoted_string> ::= any of the
128 valid 7-bit ASCII characters
(or standard abbreviations)
:TRIGger:UART:IDLE
<time_value> (see
page 577)
:TRIGger:UART:IDLE?
(see page 577)
<time_value> ::= time from 1 us
to 10 s in NR3 format
:TRIGger:UART:PARity
<parity> (see
page 578)
:TRIGger:UART:PARity?
(see page 578)
<parity> ::= {EVEN | ODD | NONE}
:TRIGger:UART:POLarit
y <polarity> (see
page 579)
:TRIGger:UART:POLarit
y? (see page 579)
<polarity> ::= {HIGH | LOW}
:TRIGger:UART:QUALifi
er <value> (see
page 580)
:TRIGger:UART:QUALifi
er? (see page 580)
<value> ::= {EQUal | NOTequal |
GREaterthan | LESSthan}
:TRIGger:UART:SOURce:
RX <source> (see
page 581)
:TRIGger:UART:SOURce:
RX? (see page 581)
<source> ::= {CHANnel<n> |
EXTernal}
<n> ::= 1-2 or 1-4 in NR1 format