Technical data
552 Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
:TRIGger:SEQuence:RESet
(see page 750)
Command Syntax
:TRIGger:SEQuence:RESet <value>
<value> ::= {NONE | PATTern1,ENTered | PATTern1,EXITed | EDGE1
| PATTern1,AND,EDGE1 | PATTern2,ENTered | PATTern2,EXITed
| EDGE2 | TIMer}
Values used in find and trigger stages are not available. EDGE2 is
not available if EDGE2,COUNt is used in trigger stage.
The :TRIGger:SEQuence:RESet command specifies the reset stage of a
sequence trigger. In multi-level trigger specifications, you may find a
pattern, then search for another in sequence, but reset the entire search to
the beginning if another condition occurs. This command accepts three
program data parameters; you can use NONE to fill out the parameter list
(for example, "EDGE1,NONE,NONE").
PATTern1 and PATTern2 are specified with the
:TRIGger:SEQuence:PATTern command. EDGE1 and EDGE2 are specified
with the :TRIGger:SEQuence:EDGE command. TIMer is specified with the
:TRIGger:SEQuence:TIMer command.
Query Syntax
:TRIGger:SEQuence:RESet?
The :TRIGger:SEQuence:RESet? query returns the reset stage specification
for a sequence trigger. NONE is returned for unused parameters.
Return Format
<reset_value><NL>
<reset_value> ::= {NONE,NONE,NONE | PATT1,ENT,NONE | PATT1,EXIT,NONE
| EDGE1,NONE,NONE | PATT1,AND,EDGE1 | PATT2,ENT,NONE
| PATT2,EXIT,NONE | EDGE2,NONE,NONE | TIM,NONE,NONE}
See Also • "Introduction to :TRIGger Commands" on page 440
• ":TRIGger:SEQuence:PATTern" on page 551
• ":TRIGger:SEQuence:EDGE" on page 549
• ":TRIGger:SEQuence:TIMer" on page 553
• ":TRIGger:SEQuence:FIND" on page 550
• ":TRIGger:SEQuence:TRIGger" on page 554