Technical data
Commands by Subsystem 5
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide 539
:TRIGger:LIN:TRIGger
(see page 750)
Command Syntax
:TRIGger:LIN:TRIGger <condition>
<condition> ::= {SYNCbreak | ID | DATA}
The :TRIGger:LIN:TRIGger command sets the LIN trigger condition to be:
• SYNCbreak — Sync Break.
• ID — Frame ID.
Use the :TRIGger:LIN:ID command to specify the frame ID.
• DATA — Frame ID and Data.
Use the :TRIGger:LIN:ID command to specify the frame ID.
Use the :TRIGger:LIN:PATTern:DATA:LENGth and
:TRIGger:LIN:PATTern:DATA commands to specify the data string length
and value.
Query Syntax
:TRIGger:LIN:TRIGger?
The :TRIGger:LIN:TRIGger? query returns the current LIN trigger value.
Return Format
<condition><NL>
<condition> ::= {SYNC | ID | DATA}
Errors • "- 241, Hardware missing" on page 709
See Also • "Introduction to :TRIGger Commands" on page 440
• ":TRIGger:MODE" on page 447
• ":TRIGger:LIN:ID" on page 529
• ":TRIGger:LIN:PATTern:DATA:LENGth" on page 532
• ":TRIGger:LIN:PATTern:DATA" on page 530
• ":TRIGger:LIN:SIGNal:DEFinition" on page 705
• ":TRIGger:LIN:SOURce" on page 536
NOTE
The ID and DATA options are available when the automotive CAN and LIN serial decode
option (Option AMS) has been licensed.