Technical data

Commands by Subsystem 5
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide 485
:TRIGger:FLEXray:EVENt:TYPE
(see page 750)
Command Syntax
:TRIGger:FLEXray:EVENt:TYPE <event>
<event> ::= {WAKeup | TSS | {FES | DTS} | BSS}
Selects the FlexRay event to trigger on. The event setting is only valid
when the FlexRay trigger mode is set to EVENt.
WAKeup — triggers on Wake- Up event.
TSS — triggers on Transmission Start Sequence event.
FES — triggers on Frame End Sequence event.
DTS — triggers on Dynamic Trailing Sequence event.
BSS — triggers on Byte Start Sequence event.
Query Syntax
:TRIGger:FLEXray:EVENt:TYPE?
The :TRIGger:FLEXray:EVENt:TYPE? query returns the currently selected
FLEXray event.
Return Format
<event><NL>
<event> ::= {WAK | TSS | {FES | DTS} | BSS}
See Also "Introduction to :TRIGger Commands" on page 440
":TRIGger:FLEXray:TRIGger" on page 491
":TRIGger:FLEXray:AUTosetup" on page 481
":TRIGger:FLEXray:SOURce" on page 490
NOTE
FES and DTS are equivalent.
NOTE
This command is only valid when the FLEXray triggering and serial decode option (Option
FLX) has been licensed.