Technical data
Commands by Subsystem 5
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide 481
:TRIGger:FLEXray:AUTosetup
(see page 750)
Command Syntax
:TRIGger:FLEXray:AUTosetup
The :TRIGger:FLEXray:AUTosetup command automatically configures
oscilloscope settings to facilitate FlexRay triggering and serial decode.
• Sets the selected source channel's impedance to 50 Ohms.
• Sets the selected source channel' s probe attenuation to 10:1.
• Sets the trigger level (on the selected source channel) to - 300 mV.
• Turns on trigger Noise Reject.
• Turns on Serial Decode.
• Sets the trigger type to FlexRay.
See Also • "Introduction to :TRIGger Commands" on page 440
• ":TRIGger:FLEXray:TRIGger" on page 491
• ":TRIGger:FLEXray:BAUDrate" on page 482
• ":TRIGger[:EDGE]:LEVel" on page 476
• ":TRIGger:FLEXray:SOURce" on page 490
NOTE
This command is only valid when the FLEXray triggering and serial decode option (Option
FLX) has been licensed.