:TRIGger:FLEXray:AUTosetup
(see page 750)
Command Syntax
:TRIGger:FLEXray:AUTosetup
The :TRIGger:FLEXray:AUTosetup command automatically configures
oscilloscope settings to facilitate FlexRay triggering and serial decode.
Sets the selected source channel's impedance to 50 Ohms.
Sets the selected source channel' s probe attenuation to 10:1.
Sets the trigger level (on the selected source channel) to - 300 mV.
Turns on trigger Noise Reject.
Turns on Serial Decode.
Sets the trigger type to FlexRay.
See Also "Introduction to :TRIGger Commands" on page 440
":TRIGger:FLEXray:TRIGger" on page 491
":TRIGger:FLEXray:BAUDrate" on page 482
":TRIGger[:EDGE]:LEVel" on page 476
":TRIGger:FLEXray:SOURce" on page 490
NOTE
This command is only valid when the FLEXray triggering and serial decode option (Option
FLX) has been licensed.