Technical data
Commands by Subsystem 5
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide 435
:TIMebase:SCALe
(see page 750)
Command Syntax
:TIMebase:SCALe <scale_value>
<scale_value> ::= 1 ns through 50 s in NR3 format
The :TIMebase:SCALe command sets the horizontal scale or units per
division for the main window.
Query Syntax
:TIMebase:SCALe?
The :TIMebase:SCALe? query returns the current horizontal scale setting in
seconds per division for the main window.
Return Format
<scale_value><NL>
<scale_value> ::= 1 ns through 50 s in NR3 format
See Also • "Introduction to :TIMebase Commands" on page 429
• ":TIMebase:RANGe" on page 433
• ":TIMebase:WINDow:SCALe" on page 439
• ":TIMebase:WINDow:RANGe" on page 438