Technical data
394 Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
:SBUS Commands
Control oscilloscope functions associated with the serial decode bus. See
"Introduction to :SBUS Commands" on page 395.
Table 66 :SBUS Commands Summary
Command Query Options and Query Returns
n/a :SBUS:CAN:COUNt:ERRor
? (see page 396)
<frame_count> ::= integer in NR1
format
n/a :SBUS:CAN:COUNt:OVERl
oad? (see page 397)
<frame_count> ::= integer in NR1
format
:SBUS:CAN:COUNt:RESet
(see page 398)
n/a n/a
n/a :SBUS:CAN:COUNt:TOTal
? (see page 399)
<frame_count> ::= integer in NR1
format
n/a :SBUS:CAN:COUNt:UTILi
zation? (see
page 400)
<percent> ::= floating-point in
NR3 format
:SBUS:DISPlay {{0 |
OFF} | {1 | ON}} (see
page 401)
:SBUS:DISPlay? (see
page 401)
{0 | 1}
n/a :SBUS:FLEXray:COUNt:N
ULL? (see page 402)
<frame_count> ::= integer in NR1
format
:SBUS:FLEXray:COUNt:R
ESet (see page 403)
n/a n/a
n/a :SBUS:FLEXray:COUNt:S
YNC? (see page 404)
<frame_count> ::= integer in NR1
format
n/a :SBUS:FLEXray:COUNt:T
OTal? (see page 405)
<frame_count> ::= integer in NR1
format
:SBUS:I2S:BASE <base>
(see page 406)
:SBUS:I2S:BASE? (see
page 406)
<base> ::= {DECimal | HEX}
:SBUS:IIC:ASIZe
<size> (see page 407)
:SBUS:IIC:ASIZe? (see
page 407)
<size> ::= {BIT7 | BIT8}
:SBUS:LIN:PARity {{0
| OFF} | {1 | ON}}
(see page 408)
:SBUS:LIN:PARity?
(see page 408)
{0 | 1}
:SBUS:M1553:BASE
<base> (see page 409)
:SBUS:M1553:BASE?
(see page 409)
<base> ::= {DECimal | HEX}
:SBUS:MODE <mode>
(see page 410)
:SBUS:MODE? (see
page 410
)
<mode> ::= {CAN | I2S | IIC | LIN
| SPI | UART}