Technical data

Commands by Subsystem 5
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide 347
:MTESt:COUNt:FWAVeforms
(see page 750)
Query Syntax
:MTESt:COUNt:FWAVeforms? [CHANnel<n>]
<n>::={1|2|3|4}forthefour channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The :MTESt:COUNt:FWAVeforms? query returns the total number of failed
waveforms in the current mask test run. This count is for all regions and
all waveforms.
Return Format
<failed><NL>
<failed> ::= number of failed waveforms in NR1 format.
See Also "Introduction to :MTESt Commands" on page 339
":MTESt:COUNt:WAVeforms" on page 350
":MTESt:COUNt:TIME" on page 349
":MTESt:COUNt:RESet" on page 348
":MTESt:SOURce" on page 369
Example Code "Example Code" on page 339