Technical data
268 Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
5 Commands by Subsystem
:HARDcopy:PALette
(see page 750)
Command Syntax
:HARDcopy:PALette <palette>
<palette> ::= {COLor | GRAYscale | NONE}
The :HARDcopy:PALette command sets the hardcopy palette color.
Query Syntax
:HARDcopy:PALette?
The :HARDcopy:PALette? query returns the selected hardcopy palette color.
Return Format
<palette><NL>
<palette> ::= {COL | GRAY | NONE}
See Also • "Introduction to :HARDcopy Commands" on page 261
• ":HARDcopy:STARt" on page 270
• ":HARDcopy:FACTors" on page 264
• ":HARDcopy:LAYout" on page 267
• ":HARDcopy:FFEed" on page 265
• ":HARDcopy:INKSaver" on page 266
NOTE
If no printer is connected, NONE is the only valid parameter.