Technical data

Commands by Subsystem 5
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide 193
:CALibrate Commands
Utility commands for viewing calibration status and for starting the user
calibration procedure. See "Introduction to :CALibrate Commands" on
page 193.
Introduction to
:CALibrate
Commands
The CALibrate subsystem provides utility commands for:
Determining the state of the calibration factor protection switch
(CAL PROTECT).
Saving and querying the calibration label string.
Reporting the calibration time and date.
Reporting changes in the temperature since the last calibration.
Table 54 :CALibrate Commands Summary
Command Query Options and Query Returns
n/a :CALibrate:DATE? (see
page 195)
<return value> ::=
<day>,<month>,<year>; all in NR1
format
:CALibrate:LABel
<string> (see
page 196)
:CALibrate:LABel?
(see page 196)
<string> ::= quoted ASCII string
up to 32 characters
:CALibrate:OUTPut
<signal> (see
page 197)
:CALibrate:OUTPut?
(see page 197)
<signal> ::= {TRIGgers | SOURce |
DSOurce | MASK}
:CALibrate:STARt (see
page 198)
n/a n/a
n/a :CALibrate:STATus?
(see page 199)
<return value> ::=
ALL,<status_code>,<status_string
>
<status_code> ::= an integer
status code
<status_string> ::= an ASCII
status string
n/a :CALibrate:SWITch?
(see page 200)
{PROTected | UNPRotected}
n/a :CALibrate:TEMPeratur
e? (see page 201)
<return value> ::= degrees C
delta since last cal in NR3
format
n/a :CALibrate:TIME? (see
page 202)
<return value> ::=
<hours>,<minutes>,<seconds>; all
in NR1 format