Technical data

Commands by Subsystem 5
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide 161
:OPERegister:CONDition (Operation Status Condition Register)
(see page 750)
Query Syntax
:OPERegister:CONDition?
The :OPERegister:CONDition? query returns the integer value contained in
the Operation Status Condition Register.
Wa i t
Trig
Run
Arm
Reg
AER?
Run bit set if oscilloscope not stopped
:OPEReration:CONDition?
Operation Status Condition Register
:OPEE
:OPEE?
Operation Status Enable
(Mask) Register
From Overload
Event Registers
To OPER bit in
Status Byte
Register
OVLR
+
OR
1023456798101112131415
Wa i t
Trig
Run
:OPERation[:EVENt]?
Operation Status Event Register
OVLR
HWE
11
5 3
From Hardware
Event Registers
HWE
12
From Mask Test
Event Registers
MTE
MTE
9
Table 49 Operation Status Condition Register
Bit Name Description When Set (1 = High = True), Indicates:
15-13 --- --- (Not used.)
12 HWE Hardware Event A hardware event has occurred..
11 OVLR Overload A 50 input overload has occurred.
10 --- --- (Not used.)
9 MTE Mask Test Event A mask test event has occurred.
8-6 --- --- (Not used.)
5Wait
Trig
Wait Trig The trigger is armed (set by the Trigger Armed
Event Register (TER)).
4 --- --- (Not used.)