Technical data

Commands by Subsystem 5
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide 129
*SRE (Service Request Enable)
(see page 750)
Command Syntax
*SRE <mask>
<mask> ::= integer with values defined in the following table.
The *SRE command sets the bits in the Service Request Enable Register.
The Service Request Enable Register contains a mask value for the bits to
be enabled in the Status Byte Register. A one in the Service Request
Enable Register enables the corresponding bit in the Status Byte Register.
A zero disables the bit.
Table 40 Service Request Enable Register (SRE)
Bit Name Description When Set (1 = High = True), Enables:
7 OPER Operation Status Register Interrupts when enabled conditions in the
Operation Status Register (OPER) occur.
6 --- --- (Not used.)
From
Operation
Status
Registers
10234567
+
OPER TRGMAVESB
RQS/
MSS
TRG
Reg
TER? Trigger Event Register
*SRE
*SRE?
*STB? Status Byte Register
Service Request Enable
(Mask) Register
Output
Queue
SRQ
From
Standard
Event
Status
Registers
USR
MSG
Service Request
OR