Technical data

104 Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Guide
4 Commands Quick Reference
n/a :WAVeform:PREamble?
(see page 606)
<preamble_block> ::= <format
NR1>, <type NR1>,<points
NR1>,<count NR1>, <xincrement
NR3>, <xorigin NR3>, <xreference
NR1>,<yincrement NR3>, <yorigin
NR3>, <yreference NR1>
<format> ::= an integer in NR1
format:
• 0 for BYTE format
• 1 for WORD format
• 2 for ASCii format
<type> ::= an integer in NR1
format:
• 0 for NORMal type
• 1 for PEAK detect type
• 2 for AVERage type
• 3 for HRESolution type
<count> ::= Average count, or 1
if PEAK detect type or NORMal; an
integer in NR1 format
n/a :WAVeform:SEGMented:C
OUNt? (see page 609)
<count> ::= an integer from 2 to
250 in NR1 format (with Option
SGM)
n/a :WAVeform:SEGMented:T
TAG? (see page 610)
<time_tag> ::= in NR3 format
(with Option SGM)
:WAVeform:SOURce
<source> (see
page 611)
:WAVeform:SOURce?
(see page 611)
<source> ::= {CHANnel<n> |
FUNCtion | MATH}
<n> ::= 1-2 or 1-4 in NR1 format
:WAVeform:SOURce:SUBS
ource <subsource>
(see page 615)
:WAVeform:SOURce:SUBS
ource? (see page 615)
<subsource> ::= {{NONE | RX} |
TX}
n/a :WAVeform:TYPE? (see
page 616)
<return_mode> ::= {NORM | PEAK |
AVER | HRES}
:WAVeform:UNSigned
{{0 | OFF} | {1 |
ON}} (see page 617)
:WAVeform:UNSigned?
(see page 617)
{0 | 1}
:WAVeform:VIEW <view>
(see page 618)
:WAVeform:VIEW? (see
page 618)
<view> ::= {MAIN}
n/a :WAVeform:XINCrement?
(see page 619)
<return_value> ::= x-increment
in the current preamble in NR3
format
Table 36 :WAVeform Commands Summary (continued)
Command Query Options and Query Returns