User`s guide
4 Triggering
170 InfiniiVision Oscilloscopes User’s Guide
15 If you choose a trigger condition that includes “Data” in its description (for example:
Rx Data) then press the Data is softkey and choose an equality qualifier. You can
choose equal to, not equal to, less than, or greater than a specific data value.
16 Use the Data softkey to choose the data value for your trigger comparison. This works
in conjunction with the Data is softkey.
17 Optional: The Burst softkey lets you trigger on the Nth frame (1-4096) after an idle
time you select. All trigger conditions must be met for the trigger to occur.
18 If Burst is selected, an idle time (1 s to 10 s) can be specified so that the oscilloscope
will look for a trigger condition only after the idle time has past. Press the Idle softkey
and rotate the Entry knob to set an idle time.
19 Remember to set the trigger mode to Normal (press
[Mode/Coupling]&Mode&Normal) to prevent Auto Triggering.
NOTE
To display UART/RS232 serial decode, see page 325.