User`s guide
4 Triggering
114 InfiniiVision Oscilloscopes User’s Guide
4 When Static or Dynamic Segments are selected:
a Press the Slot # softkey, and use the Entry knob to select the slot number to trigger
on. The maximum static slot value and the minimum dynamic slot value are limited
by the downloaded FIBEX file schedule.
b Press the Slot Type softkey to select the slot type:
• All Slots — the oscilloscope triggers on slots regardless of whether they
contain a frame.
• Empty Slots — the oscilloscope triggers when a slot does not contain a frame.
Usually used to determine if an assigned slot is empty when it should have
contained a frame.
5 Press the Cycle Rep softkey, and use the Entry knob to select the cycle repetition
factor from 2, 4, 8, 16, 32, 64, or All.
6 Press the Cycle Base softkey, and use the Entry knob to select the cycle base factor
from 0 through the Cycle Rep value minus 1.
Triggering on FlexRay Errors
1 In the Trigger section on the front panel, press the [More] key.
2 Press the Trigger softkey, and use the Entry knob to select FlexRay.
3 Press the Settings softkey to access the FlexRay Trigger Menu.
4 In the FlexRay Trigger Menu, press the Trigg er softkey, and use the Entry knob to
select Error.
NOTE
Since up to 64 cycles can be defined for FlexRay it may be helpful to set the oscilloscope
to Normal trigger mode instead of Auto trigger mode. This prevents the oscilloscope from
Auto triggering while waiting for a particular slot in a particular cycle.
Trigger
condition
VPT1000
Menu
Return to
previous menu
Select error
type