User`s guide
Triggering 4
InfiniiVision Oscilloscopes User’s Guide 113
Triggering on the FlexRay Time Schedule
In the Time trigger mode, the oscilloscope triggers on scheduled slot numbers and
segments, not actual frame contents.
1 In the FlexRay Trigger Menu press the Trigger softkey, and use the Entry knob to select
Time.
2 Press the Time softkey to access the FlexRay Time Trigger Menu.
3 Press the Segment softkey to select the segment type:
• Static Segments
• Dynamic Segments
• Symbol Segments
• Idle Segments
NOTE
To establish time triggering, the oscilloscope and VPT1000 must be set to the
synchronous mode of operation. (Press the VPT1000 softkey in the FlexRay trigger or
FlexRay Decode Menu, then press the Mode softkey. A FIBEX file that defines the
schedule must be loaded.)
Trigger
condition
VPT1000
Menu
Return to
previous menu
Open Time
Trigger Menu
Cycle base
factor
Cycle repetition
factor
Return to
previous menu
Segment type
Slot number
Slot type