User`s guide
Triggering 4
InfiniiVision Oscilloscopes User’s Guide 111
Triggering on FlexRay Frames, Times, Errors, or Events
1 To trigger on FlexRay data, you must first set up communication between the
oscilloscope and the VPT1000. See “Setting Up the Oscilloscope and the
VPT1000” on page 107.
2 In the Trigger section on the front panel, press the [More] key.
3 Press the Trigger softkey, and use the Entry knob to select FlexRay.
4 Press the Settings softkey to access the FlexRay Trigger Menu.
Triggering on FlexRay Frames
1 In the FlexRay Trigger Menu press the Trigger softkey, and use the Entry knob to select
Frame.
2 Press the Frames softkey to access the FlexRay Frame Trigger Menu.
NOTE
Before powering-off the oscilloscope, it is recommended that you change the VPT1000
mode to PC.
If oscilloscope is shut off while in Asynchronous or Synchronous modes, communication
with the VPT1000 will not be automatically restored up on power up. To restore
communication with the VPT1000, cycle the VPT1000 power, wait for its State indicator
to turn from red to green, and then press the Mode softkey (in the VPT1000 Menu) on the
oscilloscope to cycle through the modes.
FlexRay
trigger
Trigger
condition
VPT1000
Menu
Return to
previous menu
Open Frame
Trigger Menu