User`s manual

26
Instrument
Dimensions mm Inches
Width 380.6 14.98
Height 204.4 8.05
Depth 141.5 5.57
Weight kg lb
Instrument only 3.85 8.5
With accessories 4.08 9.0
Instrument shipping –
package dimensions mm Inches
Width 450 17.7
Height 250 9.84
Depth 360 14.17
Rack mount mm Inches
Width 481.6 18.961
Height 221.5 8.72
Depth 189.34 7.454
Environmental
Characteristic
Power line consumption 100 watts
Temperature Operating: 0 to +55 °C
Nonoperating: -30 to +71 °C
Humidity Operating: Up to 80% RH at or below +40 °C; up to 45% RH up to +50 °C
Non-operating: Up to 95% RH up to 40 °C; up to 45% RH up to 50 °C
Altitude Operating and non-operating: up to 4,000 m
Electromagnetic compatibility Meets EMC Directive (2004/108/EC), meets or exceeds IEC 61326-1:2005/EN
61326-1:2006 Group 1 Class A requirement
CISPR 11/EN 55011
IEC 61000-4-2/EN 61000-4-2
IEC 61000-4-3/EN 61000-4-3
IEC 61000-4-4/EN 61000-4-4
IEC 61000-4-5/EN 61000-4-5
IEC 61000-4-6/EN 61000-4-6
IEC 61000-4-11/EN 61000-4-11
Canada: ICES-001:2004
Australia/New Zealand: AS/NZS
Safety UL61010-1 2nd edition, CAN/CSA22.2 No. 61010-1-04
Vibration Meets IEC60068-2-6 and MIL-PRF-28800; class 3 random
Shock Meets IEC 60068-2-27 and MIL-PRF-28800; class 3 random;(operating 30g,
½ sine. 11 ms duration, 3 shocks/axis along major axis, total of 18 shocks
Oscilloscopes redefined: Breakthrough technology delivers more scope for the same budget
InfiniiVision X-Series physical characteristics