User`s manual

25
Oscilloscopes redefined: Breakthrough technology delivers more scope for the same budget
Performance characteristics
* Gaussian Pulse: 4 Vpp maximum into Hi-Z; 2 Vpp maximum into 50 ohms.
** Sinc, Cardiac and Gaussian Pulse: ±1.25 V into Hi-Z; +- 625 mV into 50ohms
*** Full resolution is not available at output due to internal attenuator stepping.
WaveGen – Built-in function generator
Ramp/triangle wave Frequency range: 0.1 Hz to 200 kHz
• Linearity: 1%
Variable symmetry: 0 to 100%
Symmetry resolution: 1%
Noise Bandwidth: 20 MHz typical
Sine Cardinal (Sinc) Frequency range: 0.1 Hz to 1.0 MHz
Exponential Rise/Fall Frequency range: 0.1 Hz to 5.0 MHz
Cardiac Frequency range: 0.1 Hz to 200.0 kHz
Gaussian Pulse Frequency range: 0.1 Hz to 5.0 MHz
Arbitrary Waveform length: 1 to 8k points
Amplitude Resolution: 10 bits (including sign bit)***
Repetition Rate: 0.1Hz to 12MHz
Sample Rate: 100 MSa/s
Filter Bandwidth: 20 MHz
Frequency Sine wave and ramp accuracy:
130 ppm (frequency < 10 kHz)
50 ppm (frequency > 10 kHz)
Square wave and pulse accuracy:
[50+frequency/200] ppm (frequency < 25 kHz)
50 ppm (frequency ≥ 25 kHz)
Resolution: 0.1 Hz or 4 digits, whichever is larger
Amplitude • Range:
20 mVpp to 5 Vpp into Hi-Z**
10 mVpp to 2.5 Vpp into 50 ohms**
Resolution: 100 μV or 3 digits, whichever is higher
Accuracy: 2% (frequency = 1 kHz)
DC offset Range:
±2.5 V into Hi-Z**
±1.25 V into 50 ohms**
Resolution: 100 μV or 3 digits, whichever is higher
Accuracy: ±1.5% of offset setting ±1.5% of amplitude ±1 mV
Trigger output Trigger output available on Trig out BNC
Main Output • Impedance : 50 ohms typical
• Isolation: not available, main output BNC is grounded
• Protection: overload automatically disables output