User`s manual

19
DSOX3000 Series (digital signal oscilloscope)
MSOX3000 Series (mixed signal oscilloscope)
DSOX3012A
MSOX3012A
DSOX3014A
MSOX3014A
DSOX3024A
MSOX3024A
DSOX3032A
MSOX3032A
DSOX3034A
MSOX3034A
DSOX3052A
MSOX3052A
DSOX3054A
MSOX3054A
Characteristics
Analog
bandwidth* (-3dB)
100 MHz 100 MHz 200 MHz 350 MHz 350 MHz 500 MHz 500 MHz
Analog input
channels
2442424
Calculated rise
time
≤ 3.5 ns ≤ 3.5 ns ≤ 1.75 ns ≤ 1 ns ≤ 1 ns ≤ 800 ps ≤ 800 ps
All 3000 X-Series models
Hardware bandwidth limits 20 MHz selectable
Input coupling AC, DC
Input impedance Selectable: 1 MΩ ± 1%; II 14 pF, 50 Ω ± 1.5%
Input sensitivity range 1 mV/div to 5 V/div**
Maximum sample rate 2 GSa/s per channel, 4 GSa/s half-channel interleaved
Maximum memory depth
(record length)
1 Mpt per channel, 2 Mpts half-channel interleaved (standard),
2 Mpts per channel, 4 Mpts half-channel interleaved (optional with DSOX3MEMUP)
Display 8.5-inch WVGA with 64 levels of intensity grading
Waveform update rate (max) 1,000,000 waveforms/s
Vertical resolution 8 bits
Horizontal resolution 2.5 ps
Maximum input voltage CAT I 300 Vrms, 400 Vpk; transient overvoltage 1.6 kVpk
CAT II 300 Vrms, 400 Vpk
With N2862A, N2863A or N2890A 10:1 probe: 300 Vrms
DC vertical accuracy ±[DC vertical gain accuracy + DC vertical offset accuracy + 0.25% full scale] **
DC vertical gain accuracy*
±2% full scale**
Channel-to-channel isolation > 100:1 from DC to maximum specified bandwidth of each model
(measured with same V/div and coupling on channels)
Offset range
±2 V (1 mV/div to 200 mV/div)
±50 V (> 200 mV/div to 5 V/div)
DC vertical offset accuracy ±0.1div ± 2mV ± 1% of offset setting
Oscilloscopes redefined: Breakthrough technology delivers more scope for the same budget
Performance characteristics
* Denotes warranted specifications, all others are typical.
Specifications are valid after a 30-minute warm-up period and ±10 °C from firmware calibration temperature.
** 1 mV/div and 2 mV/div are a magnification of 4 mV/div setting. For vertical accuracy calculations, use full scale of 32 mV for 1 mV/div and
2 mV/div sensitivity setting.