Technical data

Table Of Contents
:WGEN Commands 30
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 755
:WGEN:VOLTage
(see page 860)
Command Syntax
:WGEN:VOLTage <amplitude>
<amplitude> ::= amplitude in volts in NR3 format
For Sine, Square, Ramp, Pulse, and Noise waveforms, the :WGEN:VOLTage
command specifies the waveform's amplitude. Use the
:WGEN:VOLTage:OFFSet command to specify the offset voltage.
You can also specify the amplitude and offset indirectly using the
:WGEN:VOLTage:HIGH and :WGEN:VOLTage:LOW commands. For example,
an amplitude of 5 V and an offset of 1 V is the same as a high- level
voltage of 4 V and a low-level voltage of - 1 V.
Query Syntax
:WGEN:VOLTage?
The :WGEN:VOLTage? query returns the currently specified waveform
amplitude.
Return Format
<amplitude><NL>
<amplitude> ::= amplitude in volts in NR3 format
See Also "Introduction to :WGEN Commands" on page 744
":WGEN:FUNCtion" on page 746
":WGEN:VOLTage:OFFSet" on page 758
":WGEN:VOLTage:HIGH" on page 756
":WGEN:VOLTage:LOW" on page 757