Technical data

Table Of Contents
:WGEN Commands 30
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 745
:WGEN:FREQuency
(see page 860)
Command Syntax
:WGEN:FREQuency <frequency>
<frequency> ::= frequency in Hz in NR3 format
For Sine, Square, Ramp, and Pulse waveforms, the :WGEN:FREQuency
command specifies the frequency of the waveform.
You can also specify the frequency indirectly using the :WGEN:PERiod
command.
Query Syntax
:WGEN:FREQuency?
The :WGEN:FREQuency? query returns the currently set waveform
generator frequency.
Return Format
<frequency><NL>
<frequency> ::= frequency in Hz in NR3 format
See Also "Introduction to :WGEN Commands" on page 744
":WGEN:FUNCtion" on page 746
":WGEN:PERiod" on page 753