Technical data

Table Of Contents
682 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
28 :TRIGger Commands
:TRIGger:RUNT:QUALifier
(see page 860)
Command Syntax
:TRIGger:RUNT:QUALifier <qualifier>
<qualifier> ::= {GREaterthan | LESSthan | NONE}
The :TRIGger:RUNT:QUALifier command selects the qualifier used for
specifying runt pulse widths:
GREaterthan — triggers on runt pulses whose width is greater than the
:TRIGger:RUNT:TIME.
LESSthan — triggers on runt pulses whose width is less than the
:TRIGger:RUNT:TIME.
NONE triggers on runt pulses of any width.
Query Syntax
:TRIGger:RUNT:QUALifier?
The :TRIGger:RUNT:QUALifier? query returns the runt trigger qualifier
setting.
Return Format
<qualifier><NL>
<qualifier> ::= {GRE | LESS NONE}
See Also "Introduction to :TRIGger Commands" on page 641
":TRIGger:MODE" on page 648
":TRIGger:RUNT:TIME" on page 684