Technical data

Table Of Contents
638 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
27 :TIMebase Commands
:TIMebase:WINDow:RANGe
(see page 860)
Command Syntax
:TIMebase:WINDow:RANGe <range value>
<range value> ::= range value in seconds in NR3 format
The :TIMebase:WINDow:RANGe command sets the full- scale horizontal time
in seconds for the zoomed (delayed) window. The range is 10 times the
current zoomed view window seconds per division setting. The main sweep
range determines the range for this command. The maximum value is one
half of the :TIMebase:RANGe value.
Query Syntax
:TIMebase:WINDow:RANGe?
The :TIMebase:WINDow:RANGe? query returns the current window
timebase range setting.
Return Format
<value><NL>
<value> ::= range value in seconds
See Also "Introduction to :TIMebase Commands" on page 630
":TIMebase:RANGe" on page 633
":TIMebase:POSition" on page 632
":TIMebase:SCALe" on page 635