Technical data

Table Of Contents
508 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
24 :SBUS<n> Commands
:SBUS<n>:LIN:SIGNal:BAUDrate
(see page 860)
Command Syntax
:SBUS<n>:LIN:SIGNal:BAUDrate <baudrate>
<baudrate> ::= integer from 2400 to 625000 in 100 b/s increments
The :SBUS<n>:LIN:SIGNal:BAUDrate command sets the standard baud rate
of the LIN signal from 2400 b/s to 625 kb/s in 100 b/s increments. If you
enter a baud rate that is not divisible by 100 b/s, the baud rate is set to
the nearest baud rate divisible by 100 b/s.
Query Syntax
:SBUS<n>:LIN:SIGNal:BAUDrate?
The :SBUS<n>:LIN:SIGNal:BAUDrate? query returns the current LIN baud
rate setting.
Return Format
<baudrate><NL>
<baudrate> ::= integer from 2400 to 625000 in 100 b/s increments
See Also "Introduction to :TRIGger Commands" on page 641
":TRIGger:MODE" on page 648
":SBUS<n>:LIN:TRIGger" on page 512
":SBUS<n>:LIN:SIGNal:DEFinition" on page 816
":SBUS<n>:LIN:SOURce" on page 509