23 :SAVE Commands
:SAVE:WMEMory:SOURce
(see page 860)
Command Syntax
:SAVE:WMEMory:SOURce <source>
<source> ::= {CHANnel<n> | FUNCtion | MATH | WMEMory<r>}
<n> ::= 1 to (# analog channels) in NR1 format
<r> ::= {1 | 2}
The :SAVE:WMEMory:SOURce command selects the source to be saved as a
reference waveform file.
Query Syntax
:SAVE:WMEMory:SOURce?
The :SAVE:WMEMory:SOURce? query returns the source to be saved as a
reference waveform file.
Return Format
<source><NL>
<source> ::= {CHAN<n> | FUNC | WMEM<r> | NONE}
See Also "Introduction to :SAVE Commands" on page 434
":SAVE:WMEMory[:STARt]" on page 451
":RECall:WMEMory<r>[:STARt]" on page 431
NOTE
Only ADD or SUBtract math operations can be saved as reference waveforms.
NOTE
MATH is an alias for FUNCtion. The query will return FUNC if the source is FUNCtion or
MATH.