Technical data

Table Of Contents
:MTESt Commands 20
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 409
:MTESt:RMODe:SIGMa
(see page 860)
Command Syntax
:MTESt:RMODe:SIGMa <level>
<level> ::= from 0.1 to 9.3 in NR3 format
When the :MTESt:RMODe command is set to SIGMa, the
:MTESt:RMODe:SIGMa command sets the test sigma level to which a mask
test runs. Test sigma is the best achievable process sigma, assuming no
failures. (Process sigma is calculated using the number of failures per
test.) The test sigma level indirectly specifies the number of waveforms
that must be tested (in order to reach the sigma level).
Query Syntax
:MTESt:RMODe:SIGMa?
The :MTESt:RMODe:SIGMa? query returns the current Sigma level setting.
Return Format
<level><NL>
<level> ::= from 0.1 to 9.3 in NR3 format
See Also "Introduction to :MTESt Commands" on page 387
":MTESt:RMODe" on page 404
Example Code "Example Code" on page 387