Technical data

Table Of Contents
:MTESt Commands 20
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 391
:MTESt:AMASk:CREate
(see page 860)
Command Syntax
:MTESt:AMASk:CREate
The :MTESt:AMASk:CREate command automatically constructs a mask
around the current selected channel, using the tolerance parameters
defined by the :MTESt:AMASk:XDELta, :MTESt:AMASk:YDELta, and
:MTESt:AMASk:UNITs commands. The mask only encompasses the portion
of the waveform visible on the display, so you must ensure that the
waveform is acquired and displayed consistently to obtain repeatable
results.
The :MTESt:SOURce command selects the channel and should be set before
using this command.
See Also "Introduction to :MTESt Commands" on page 387
":MTESt:AMASk:XDELta" on page 394
":MTESt:AMASk:YDELta" on page 395
":MTESt:AMASk:UNITs" on page 393
":MTESt:AMASk:SOURce" on page 392
":MTESt:SOURce" on page 417
Example Code "Example Code" on page 387