Technical data

Table Of Contents
276 Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
15 :FUNCtion Commands
:FUNCtion:GOFT:SOURce2
(see page 860)
Command Syntax
:FUNCtion:GOFT:SOURce2 <value>
<value> ::= CHANnel<n>
<n> ::= {{1 | 2} | {3 | 4}} for 4ch models, depending on SOURce1
selection
<n> ::= {1 | 2} for 2ch models
The :FUNCtion:GOFT:SOURce2 command selects the second input channel
for the g(t) source that can be used as the input to the FFT, INTegrate,
DIFF (differentiate), or SQRT functions.
If CHANnel1 or CHANnel2 is selected for :FUNCtion:GOFT:SOURce1, the
SOURce2 selection can be CHANnel1 or CHANnel2. Likewise, if CHANnel3
or CHANnel4 is selected for :FUNCtion:GOFT:SOURce1, the SOURce2
selection can be CHANnel3 or CHANnel4.
Query Syntax
:FUNCtion:GOFT:SOURce2?
The :FUNCtion:GOFT:SOURce2? query returns the current selection for the
second input channel for the g(t) source.
Return Format
<value><NL>
<value> ::= CHAN<n>
<n> ::= {{1 | 2} | {3 | 4}} for 4ch models, depending on SOURce1
selection
<n> ::= {1 | 2} for 2ch models
See Also "Introduction to :FUNCtion Commands" on page 271
":FUNCtion:GOFT:SOURce1" on page 275
":FUNCtion:GOFT:OPERation" on page 274