Common (*) Commands 5
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide 137
*OPT (Option Identification)
(see page 860)
Query Syntax
*OPT?
The *OPT? query reports the options installed in the instrument. This
query returns a string that identifies the module and its software revision
level.
Return Format
0,0,<license info>
<license info> ::= <All field>, <reserved>, <MSO>, <reserved>,
<Memory>, <Low Speed Serial>, <Automotive Serial>,
<reserved>, <reserved>,
<Power Measurements>, <RS-232/UART Serial>,
<Segmented Memory>, <Mask Test>, <reserved>,
<Bandwidth>, <reserved>, <reserved>, <reserved>,
<I2S Serial>, <reserved>, <reserved>,
<Waveform Generator>, <reserved>, <reserved>
<All field> ::= {0 | All}
<reserved> ::= 0
<MSO> ::= {0 | MSO}
<Memory> ::= {0 | MEMUP}
<Low Speed Serial> ::= {0 | EMBD}
<Automotive Serial> ::= {0 | AUTO}
<Power Measurements> ::= {0 | PWR}
<RS-232/UART Serial> ::= {0 | COMP}
<Segmented Memory> ::= {0 | SGM}
<Mask Test> ::= {0 | MASK}
<Bandwidth> ::= {0 | BW20 | BW50}
<I2S Serial> ::= {0 | AUDIO}
<Waveform Generator> ::= {0 | WAVEGEN}
The <MSO> field indicates whether the unit is a mixed-signal
oscilloscope.
The *OPT? query returns the following:
See Also "Introduction to Common (*) Commands" on page 127
Module Module Id
No modules attached 0,0,0,0,MSO,0,0,0,0,0,0,0,0,0,0,0,0,0,0,0,0,0,0,0,0
,0