Specifications
32 | Keysight | InniiVision 3000T X-Series Oscilloscopes - Data Sheet
Performance characteristics
1. Gaussian Pulse: 4 Vpp maximum into Hi-Z; 2 Vpp maximum into 50 Ω.
WaveGen – Built-in function/arbitrary waveform generator (specications are typical) (continued)
Frequency Sine wave and ramp accuracy:
130 ppm (frequency < 10 kHz)
50 ppm (frequency > 10 kHz)
Square wave and pulse accuracy:
[50+frequency/200] ppm (frequency < 25 kHz)
50 ppm (frequency ≥ 25 kHz)
Resolution: 0.1 Hz or 4 digits, whichever is larger
Amplitude Range:
20 mVpp to 5 Vpp into Hi-Z
1
10 mVpp to 2.5 Vpp into 50 Ω
1
Resolution: 100 μV or 3 digits, whichever is higher
Accuracy: 2% (frequency = 1 kHz)
DC offset Range:
±2.5 V into Hi-Z
1
±1.25 V into 50 Ω
1
Resolution: 100 μV or 3 digits, whichever is higher
Accuracy (waveform modes): ± 1.5% of offset setting ± 1% of amplitude ± 1 mV
Accuracy (DC mode): ± 1.5% of offset setting ± 3 mV
Trigger output Trigger output available on Trig out BNC
Main Output Impedance : 50 Ω typical
Isolation: not available, main output BNC is grounded
Protection: overload automatically disables output
Digital voltmeter (specications are typical)
Functions ACrms, DC, DCrms
Resolution ACV/DCV: 3 digits
Measuring rate 100 times/ second
Autoranging Automatic adjustment of vertical amplication to maximize the dynamic range of measurements.
Range meter Graphical display of most recent measurement, plus extrema over the previous 3 seconds.
Precision counter / totalizer (specication are typical)
Counter Source Any analog channel or trigger qualied event
Resolution 8 digits (8 digits for trigger qualied event)
Max frequency 1 GHz
Trig qual events 1/(trigger hold off time) for trigger qualied events (max 25 MHz, ,minimum dead time of 40 ns)
Measurement Frequency, period, totalize
Totalizer Counter size 64 bit totalizing counter
Edge Rise or fall
Gating Positive or negative level. Select from analog channels except the source.
Time reference Internal or external 10 MHz reference clock