Forcing a Default Setup
If your Infiniium oscilloscope is not working properly when you start it up
and pressing [Default Setup] does not fix the problem, follow these steps to
force a default setup and return the Infiniium to normal operation.
1
Turn off the Infiniium.
2
Turn on the Infiniium. If the oscilloscope does not successfully restart
then try recycling the power again.
3
After the Windows 7 load screen disappears, press the [Default Setup]
key on the front panel. If the oscilloscope still does not successfully
restart, follow the instructions for recovering the hard drive.
Infiniium Hard Drive Recovery
Follow these steps to recover your Infiniium hard drive.
1
Turn off the Infiniium.
2
Connect the keyboard to the keyboard connector on the rear panel of
the oscilloscope.
3
Connect the mouse to the mouse connector on the rear panel of the
oscilloscope.
4
Turn on the Infiniium.
5
When you see the message press F2 to boot the Infiniium Recovery
System, press F2 and follow the on- screen instructions.
6
Once the recovery process is finished and the oscilloscope is running,
check in the About Infiniium dialog box under installed options to see
if all of the options you ordered are installed. If the options are not
installed, please install them using the license keys provided on the
oscilloscope option license certificates you received, or refer to the rear
panel of the oscilloscope.