User`s guide
16 Agilent Infiniium 90000 X-Series Oscilloscopes User’s Guide
1
Setting Up the Oscilloscope
Connecting power
1
Connect the power cord to the rear of the oscilloscope, then to a
suitable AC voltage source.
Power Supply Protection Features
The Infiniium 90000 X- Series oscilloscope’s power supply has built- in
protection features for over- voltage, over- current, and over- temperature
fault conditions.
Green LEDs for AC and DC indicate normal operation. A red Fault LED
indicates a fault condition.
When a fault occurs (and the power supply shuts down), you must unplug
the power cord, wait 10 to 30 seconds, and plug the power cord back in to
reset the protection latch and allow the oscilloscope to be powered on
again.
If the power supply has overheated, you may need to wait longer so it can
cool before you can power on the oscilloscope again.
Tabl e 3 Power Requirements
Power 100 - 240 VAC at 50/60 Hz
Input power 800 Watts
CAUTION
Use only power cords designed for your oscilloscope.
The line cord provided is matched by Agilent to the country of origin of the order.
WARNING
To avoid electric shock, be sure the oscilloscope is properly grounded.
NOTE
Note that the oscilloscope’s power supply has a built-in safety feature that, after the
oscilloscope is turned off, requires the oscilloscope to remain off for about 10 seconds
before it is turned back on.