Specifications
Chapter 5: Replacing Assemblies
Removing and Replacing RealEdge Components
92
4
Carefully remove the RealEdge module.
Figure 5-48
5
Remove the two screws on the side of the attenuator. When replacing, torque
to 5 in- lbs.
Figure 5-49
6
Loosen the SMA connector from the male SMA connector on the 1.85 connector.
When replacing, torque the connector to 8 in- lbs.
Figure 5-50
7
Remove the attenuator assembly.