Specifications
Chapter 4: Troubleshooting
Acquisition/Backplane Assembly Troubleshooting
60
Acquisition/Backplane Assembly Troubleshooting
This section describes which board assembly to replace if any of the scope self tests
fail. When the self- test error message file is generated it is sent to the following location:
C:\ProgramData\Agilent\Infiniium\selftest\SelfTestLog.txt
The error message usually indicates the channel with the error. When identifying which
acquisition assembly to replace, remember that the acquisition assembly nearest the
bottom of the oscilloscope is for channel 1. The one above it is for channel 2. The next
higher is for channel 3, and the top acquisition assembly is for channel 4. Replace the
acquisition assembly that has the error.
If the error message does not indicate a channel, refer to the following table to determine
which assembly to replace.
Test Group and Test Name Error Type Assembly to Replace
Timebase Test Group
Timebase Interpolator Test Backplane assembly
ADC Test Group
ADC Register Tests ADC1, ADC2, ADC3, ADC4 Acquisition assembly 1
ADC5, ADC6, ADC7, ADC8 Acquisition assembly 2
ADC9, ADC10, ADC11, ADC12 Acquisition assembly 3
ADC13, ADC14, ADC15, ADC16 Acquisition assembly 4
ADC Voltage Test Points ADC1, ADC2, ADC3, ADC4 Acquisition assembly 1
ADC5, ADC6, ADC7, ADC8 Acquisition assembly 2
ADC9, ADC10, ADC11, ADC12 Acquisition assembly 3
ADC13, ADC14, ADC15, ADC16 Acquisition assembly 4
Misc. Scope Test Group
Temp Sensor Tests Hedwig0, Hedwig1, Hedwig2, Hedwig3, Oak0 Acquisition assembly 1
Hedwig4, Hedwig5, Hedwig6, Hedwig7, Oak2 Acquisition assembly 2
Pred1, Pred2, MainFPGA, Wahoo, Ambient Backplane assembly
Merwig0, Merwig1 Replace acquisition assembly 1.
If error still occurs, put original
acquisition assembly back into
scope and replace backplane
assembly.
Merwig2, Merwig3 Replace acquisition assembly 2.
If error still occurs, put original
acquisition assembly back into
scope and replace backplane
assembly.
Merwig4, Merwig5, Hedwig8, Hedwig9,
Hedwig10, Hedwig11, Oak4
Replace acquisition assembly 3.
If error still occurs, put original
acquisition assembly back into
scope and replace backplane
assembly.