Specifications
Chapter 4: Troubleshooting
Power Supply Troubleshooting
55
LED Type of Fault What To Do
AOT0* Over-temperature Replace acquisition assembly 4
AOT1* Over-temperature Replace acquisition assembly 4
AOT2* Over-temperature Replace acquisition assembly 3
AOT3* Over-temperature Replace acquisition assembly 3
AOT4* Over-temperature Replace acquisition assembly 2
AOT5* Over-temperature Replace backplane assembly 2
AOT6* Over-temperature Replace backplane assembly 1
AOT7* Over-temperature Replace acquisition assembly 1
AOV0 Under-voltage Replace acquisition assembly 4
AOV1 Under-voltage Replace acquisition assembly 3
AOV2 Under-voltage Replace acquisition assembly 2
AOV3 Under-voltage Replace acquisition assembly 1
AUV0 Over-voltage Replace acquisition assembly 4
AUV1 Over-voltage Replace acquisition assembly 3
AUV2 Over-voltage Replace acquisition assembly 2
AUV3 Over-voltage Replace acquisition assembly 1
BOV0 Over-voltage Replace assembly indicated by
illuminated AUV0 or FBS LED
BOV1 Used for internal testing
purposes only
N/A
BOV2 Any Any problem that causes a
shutdown also turns on this LED
BOV3 Used for internal testing
purposes only
N/A
BOVG Over-voltage Replace backplane assembly
BSF Over-voltage Replace backplane assembly
FBS Over-voltage Replace backplane assembly
FLT Over-voltage Replace backplane assembly
*If an AOT LED is Lit
If the oscilloscope shuts down and the AOT0, AOT1, AOT2, AOT3, AOT4, AOT5, AOT6, or
AOT7 indicator LED is lit, then one of the scope channels is getting too hot or the
temperature sensing circuitry has failed. Normally, you would have to turn the
oscilloscope on and leave it on for awhile before you would see a shutdown caused by
temperature problems. If all the fans are turning, it is unlikely that the oscilloscope would
shut down due to temperature unless the airflow is being restricted or the input air is
already too hot to cool the insides of the oscilloscope.
It is possible that all the fans are working, but the oscilloscope still shuts down because
of temperature due to one or more of the channels malfunctioning and drawing too much
current. If so, you would need to replace the indicated acquisition assembly. It is also
possible for one of the heat sinks on one of the acquisition assemblies' critical parts to
become detached and cause that part to overheat and shut down the oscilloscope.