Specifications

Chapter 4: Troubleshooting
Power Supply Troubleshooting
52
Power Supply Troubleshooting
This section provides information to help you isolate the problem to the assembly level
when the power system is not operating properly.
There are two sets of LEDs on the backplane assembly. Seeing which of these LEDs
are illuminated will help you identify the reason and the remedy for the power failure.
WARNING SHOCK HAZARD!
The maintenance described in this section is performed with power supplied to the
oscilloscope and with the protective covers removed. Only trained service personnel
who are aware of the hazards involved should perform the maintenance. Read the safety
notices at the back of this guide before proceeding. Failure to observe safety precautions
may result in electric shock.
To locate both sets of LEDs, remove the oscilloscope top panel to expose the top edges
of the backplane assembly. The first set of LEDs is located in the upper right corner
of the backplane assembly, to the right of the J13001 connector (see Figure 4- 5).
Note: If you cannot see either set of LEDs, you may have a slightly modified backplane
board. Contact your Agilent Service Center for assistance.
Figure 4-5 First set of LEDs
This set of LEDs works in groups. To use them to determine the problem and remedy,
refer to the following table and follow these steps:
1 Connect the power cord. All 16 LEDs should illuminate. If none of them are illuminated,
replace the bulk supply.
2 Turn power on. If all 16 LEDs turn off, the power supplies are working.