Specifications

Chapter 3: Testing Performance
Analog Bandwidth -- Maximum Frequency Test
34
8 Measure the input power to the scope channel and convert this measurement to Volts
RMS using the expression:
For example, if the power meter reading is 4.0 μW, then Vin = (4.0*10
-6
* 50Ω)
1/2
=
14.1 mVrms.
Record the RMS voltage in the Analog Bandwidth - Maximum Frequency Check section
of the Performance Test Record (Vin @ 50 MHz).
9 Press [Clear Display] on the scope and record the scope V rms reading in the Analog
Bandwidth - Maximum Frequency Check section of the Performance Test Record (Vout
@ 50 MHz).
10 Calculate the reference gain as follows:
Record this value in the Calculated Gain @50 MHz column in the Analog Bandwidth -
Maximum Frequency Check section of the Performance Test Record.
Notes
For all scope readings in this procedure, use the mean value in the Measurements display area at the
bottom of the screen.
V
in
P
meas
50Ω×=
Record the
mean reading
Gain
50 MHz
V
out @50 MHz
V
in @50 MHz
-------------------------------
=