Specifications

Chapter 3: Testing Performance
Analog Bandwidth -- Maximum Frequency Test
33
f When the Enter Measurement Info window appears, configure this measurement
as follows:
Source = Channel 1
Measurement Area = Entire Display
RMS Type = AC
7 Set the generator to apply a 50 MHz sine wave with a peak- to- peak amplitude of about
4 divisions.
Use the following tables to determine the approximate required signal amplitude.
The amplitude values in the tables are not absolutely required. If your generator
is unable to produce the recommended amplitude, then set the generator to the
highest value that does not produce a vertically clipped signal on the scope.
Table 3-1. Nominal Generator Amplitude Settings for all Oscilloscope Models, channels 1-4
Table 3-2. Nominal Generator Amplitude Settings for 50 and 62 GHz Oscilloscope Models, channels 1R-3R
Scope
Vertical Sensitivity
Generator Signal
Amplitude (Vp-p)
Generator Signal
Amplitude (dBm)
10 mV/div 0.08 -18
20 mV/div 0.16 -12
50 mV/div 0.4 -4
100 mV/div 0.8 +2
200 mV/div 1.6 +8
500 mV/div 4.0 +16
1 V/div 6.3 +20
Scope
Vertical Sensitivity
Generator Signal
Amplitude (Vp-p)
Generator Signal
Amplitude (dBm)
10 mV/div 0.08 -12.5
20 mV/div 0.16 -9.5
50 mV/div 0.4 -5.5
100 mV/div 0.8 -2.5