Specifications

Chapter 5: Replacing Assemblies
Setting the Calibration Factors after Replacing an Acquisition Board
101
Setting the Calibration Factors after Replacing an Acquisition
Board
You must perform a user calibration and self test procedure after replacing an
acquisition board. Refer to the calibration chapter in this Service Guide for
details.
Let the oscilloscope warm up before testing.
The oscilloscope under test must be warmed up (with the oscilloscope application running) for at least
30 minutes prior to the start of any performance test.