Specifications
NOTE 
When loading a Result File for calculating Sigma vs. Tau, 
the carrier frequency used for the measurement must be 
entered in the Instrument Parameters display in order to 
compute accurate values. 
Detector/Discr. Input Frequency is the frequency of the input signal 
connected to the HP 11848A's phase detector. The Detector/Discriminator 
Input Frequency is used to: 
• Select the appropriate Low Pass Filter (LPF) in the HP 11848A Phase 
Noise Interface. For frequencies < 95 MHz, the system enables the 
2 MHz LPF to eliminate feedthrough of the fundamental signal and 
the additive products out of the phase detector. When the 2 MHz LPF 
is enabled, the maximum offset frequency the system can measure is 
2 MHz. 
• Set the frequency of the Reference Source when the Reference Source 
is under system control. 
• When using an HP 11729C under System Control, the HP 3048A will 
derive the correct Detector/Discr. Input Frequency and enter it into the 
Instrument Parameter display. If the currently entered Detector/Discr. 
Input Frequency is less than 95 MHz, and you are allowing the 
HP 3048A to input the correct Detector/Discr. Input Frequency by 
computing the value from a System Controlled HP 11729C, then the 
measured offset range will be limited to 2 MHz. To eliminate this 
problem, set the Detect/Discr. Input Frequency to a frequency greater 
than 95 MHz. 
Instr. Params: $N w/PLL 2-15 