How to Use Noise Monitor 
Evaluating a Single Offset Frequency 
1.
 Press the SPAN softkey to increment the frequency span. The increment 
values are the measurement ranges denned in the segment tables. 
2.
 Press 
Take Sweep 
Take Sweep sets up the Interface, autoranges the 
analyzer, measures the span, and holds the averaged display. 
NOTE 
Noise Monitor allows the user to manipulate the Front Panel 
controls of the analyzer. After the Interface is set up, the 
analyzer settings can be changed. 
(Note that changing the bandwidth to other than the one 
set up by the Take Sweep softkey will cause the Marker 
function to return inaccurate data.) 
3.
 Place the analyzer's marker at the desired frequency. 
4.
 Select the appropriate Marker softkey to read and display the value of 
the marker. 
Evaluating Phase Fluctuations as a Function of Time 
Real Time Noise measurements can be useful for evaluating noise tran-
sients or device sensitivity to mechanical or electrical stimulation. Contin-
uous Real Time display of the noise measurement can be made by turning 
the averaging function on the analyzer off (or by setting the analyzer for 
repeat display or continuous sweep.) 
5-32 Noise Monitor / NewNse Monitor 
rev.05SEP89 