Specifications
There are several methods available for calibrating the Phase Detector 
Constant when phase noise measurements are made without using a 
phase lock loop. The primary considerations for selecting a calibration 
method are measurement accuracy and equipment availability. 
Technique 
Use the Current Detector 
Constant 
Derive From Measured +/— DC 
Peak 
Derive From Measured 
Beat-note 
Derive From Double Sided Spur 
Derive From Single Sided Spur 
Additional Equipment 
Required 
None 
Phase Shifter or source with 
appropriate frequency resolution 
Second RF source 
Phase Modulator, Audio Source, 
and RF Analyzer 
Directional Coupler, RF 
Analyzer, and Second RF 
Source 
Accuracy 
* 
Fair 
Good 
Best 
Best 
* Accuracy depends on which technique was used to determine the Detector Constant 
in memory. 
2-58 Calibr Process: $N w/o PLL 