User`s guide

Chapter 4 101
Performance Tests
Performance Test Procedures
Performance Tests
Table 4-1 Recommended Test Equipment for 11974A
Instrument Critical Specifications Recommended
Model
Spectrum Analyzer LO and IF ranges compatible with 11974A
SWEEP + TUNE voltage compatible with
11974A.
8566B
w/serial prefix 2948
or later
Synthesized Sweeper Frequency: 13.25 to 20 GHz
External leveling capability
Sub harmonics and multiples thereof: <
25 dBc
8340B
a
a. If an 8340A is used, the 1.0 V/GHz output will have to be modified to a 0.5
V/GHz output for proper display accuracy, leveling flatness, and harmonic sup-
pression. Provided with your mm-wave source module is a set of modification
instructions. This modification procedure will enable you to properly convert
your 8340A to a 0.5 V/GHz output.
Amplifier Frequency Range: 3 to 5 GHz
Output Level: +16 dBm leveled
11975A
Amplifier 15 dB gain from 13.25 to 20 GHz
Compatible with mm-wave source module
Non-harmonic spurious: <
50 dBc
8349B
Power Meter Compatible with thermistor mount 432A
Thermistor Mount SWR: <2.0 from 26.5 to 40 GHz R486A
Power Sensor SWR: <1.2 at 5 GHz 8481A
Power Meter Compatible with power sensor 436A
Directional Coupler Coupling: 10 dB
Directivity: >40 dB
Primary line SWR: 1.04
Auxiliary arm SWR: 1.05
R752CS
Isolator Insertion Loss: <1.5 dB
Isolation: >25 dB
SWR: <1.4
R365A
Millimeter-Wave
Source Module
Output power: +2 dBm leveled
Output SWR: <2.0
Harmonically related spurious signals: <
20 dBc
83554A
Waveguide Straights 5 cm minimum
SWR: <1.016
R896B