Installation and Verification Manual HP 70903A IF Section SERIAL NUMBERS This manual applies directly to HP 70903A IF Sections with serial numbers prefixed 2835A and below. @!a HEWLETT PACKARD HP Part No.
Notice The information contained in this document is subject to change without notice. Hewlett-Packard makes no warranty of any kind with regard to this material, including, but not limited to, the implied warranties of merchantability and fitness for a particular purpose. Hewlett-Packard shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Restricted Rights Legend.
CERTIFICATION Hewlett-Packard Company certifies that this product met its published specifications at the time of shipment from the factory. Hewlett-Packard further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology (NET, formerly NBS), to the extent allowed by the institute3 calibmtion facility, and to the calibration facilities of other International Standards Organization members.
SAFETY SYMBOLS The following safety symbols are used throughout this manual and in the instrument. Familiarize yourself with each of the symbols and its meaning before operating the instrument. Note Notes will be used to cell attention to an important point in the text. This is not a hazard, but should not be ignored as it will provide a better understanding of the text. Caution The CAUTION sign denotes a hazard.
HP 70000 MODULAR MEASUREMENT SYSTEM DOCUMENTATION OUTLINE Instruments and modules of the HP 70000 Modular Measurement System are documented to varying levels of detail. Modules that serve as masters (that is, control other elements) of an instrument require operation information in addition to installation and verification instructions. Modules that function as slaves in a system require only a subset of installation and verification information.
Contents 1. General Information l- l ........ Introduction ........... l- l ........ Module Description ......... l-2 Modular Measurement System Terms . ........ l-2 ........ Functional Terms ......... l-2 . ........ Structural Terms ........ l-3 ........ Safety Considerations ........ l-3 . ........ Modules Covered by Manual .... l-3 ........ Serial Numbers .......... l-4 ........ Manual Updating Supplement .... l-4 ........ Initial Inspection .......... l-4 ........ Accessories ........... l-5 ........ .
2-3 Setting the HP-MSIB Address Switches ........ ........ 2-4 Installing the Module in the Mainframe 2-4 Module Installation .......... 2-5 Connecting the Rear-Panel Cables ........ ...... 2-5 Rear-Panel Cables for HP 71100A RF Spectrum Analyzer 2-5 HP 71100A IF Section Cable Connections ....... 2-6 Rear-Panel Cabling for HP 71210A Microwave Spectrum Analyzer .. ...... 2-6 HP 71210A Dual-IF Section Cable Connections 3. Spec&ations 4. Ver&ation 41 Introduction ............... ....... 41 ........... .
Figure l-l Typical Serial Number Label l-3 Figure l-2 Module Front- and Hear-Panel Features l-7 Figure l-3 Example of a Static-Safe Work Station 1-8 Figure l-4 Mainframe Packagiug Material 1-14 Figure l-5 Module Packaging Material 1-15 Figure 2-l Module Address Switches 2-3 Figure 2-2 Module Installation 2-4 Figure 2-3 HP 71100 Spectrum Analyzer Rear-Panel Cables 2-5 Figure 2-4 HP 71210 Spectrum Analyzer Rear-Panel Cables 2-6 Accessories Shipped with a Single Module l-4 List of
General Information Introduction The HP 70993A Installation and Verification Manual contains information needed to install and verify the HP 76903A IF Section. For information on instailing and verifying HP 70000 Modular Measurement Systems, refer to the installation and verification manual for the system master (for example, HP 7096OA Local Oscillator).
Modular Measurement System Terms Understanding the following terms is essential to understanding HP-MSIB (Hewlett-Packard Modular System Interface Bus) addressing and the structural relationships among modular system devices. Functional Terms Modular system devices may be combined to allow them to communicate and operate as an instrnment. The following terms identify the interrelationships among devices within a modular instrument.
Safety Considerations Before operating this module, familiarize yourself with any safety markings on the module and the safety instmctions in this manual. This module has been manufactured and tested according to international safety standards. The cautions and warnings in this manual must be followed to ensure the safe operation of the module and protection of personnel. Refer to the summary of safety considerations at the front of this manual.
Manual Updating Supplement A module manufactured after this manual was printed may have a serial number prefix other than that listed under “Serial Numbers ” on the manual title page. A higher serial number prefix than stated on the title page indicates that changes have been made to the module since the manual was printed. Any changes that affect information in this manual are documented in the Manual Updating Supplement for this manual.
Front/Rear-Panel Features Figure l-2 shows the HP 70903A front- and rear-panel features. Front-Panel Status Indicators It is normal for both of the front-panel status LED indicators to flash on, then off, during the module self-test. Self-test occurs each time the instrument is turned on. Listed below are other reasons why each status LED might light.
Rear-Panel Inputs and Outputs 21.4MHz IN This SMB (m) input connector receives a from the RF section. 21.4-MHz signal, at a nominal level of -5 dBm, 21.4.MHz OUT This SMB (m) output connector is used when cascading signals iu a HP 70000 Modular Measurement System that has two IF sections. VIDEO OUT This SMB (m) output connector supplies a typical 0-2V video signal to the system master (for example, HP 70900A Local Oscillator).
HP70903A IF SECTIDN RES BW 100kHz 3MHz c I I \ ‘21.4 MHz 21.4 MHz OUT \ HP70903A I F mAUX OUTPUT1V VIDEO - 0 Oo 0 ooooooooo 0 0 0 0 0 0 0 0000 IF 21.4 MHz O O O O ’o”oooooo 0000 0000000000 0 OoO @ P Figure l-2.
Electrostatic Discharge Information Electrostatic discharge (ESD) can damage or destroy electronic components. All work on electronic assemblies should be performed at a static-safe work station. Figure l-3 shows an example of a static-safe work station using two types of ESD protection: (1) conductive table-mat and wrist-strap combination, (2) conductive table-mat and heel-strap combination. The two types must be used together to ensure adequate ESD protection.
Reducing ESD Damage Handling of Electronic Components n Perform work on these items at a static-safe work station. n Store or transport these items in static-shielding containers. n Use proper handling techniques. Caution PC board traces are easily damaged. Do not touch traces with the bare hands. Always handle board assemblies by the edges.
Static-Safe Accessories Table l-2. Static-Safe Accessories Description set includes: 3M static control mat 0.6-m by 1.2-m (Zft by 4ft) and 4.6cm (15ft)ground wire (The wristcstrap and wrist-strap cord are not included. They must be ordered separately.) 9300-0980* Wrist strap cord 1.5m (5 R) 93ocbo985* Wrist strap (large) ~Wrist strap (small) ESD heel strap (reusable 6 to 12 months) ‘Shoe ground strap (one time use only) *Can be ordered through any HewlettrPackard G&e.
Table l-3. Hewlett-Packard Sales and Service Offices IN THE UNITED STATES caIifornia HewletePackard Co. 1421 South Manhattan Ave. P-0. Box 4230 Fullerton, CA 92631 (714) 999-6700 IN AUSTRALIA Hewlett-Packard Australia Ltd. 31-41 Joseph Street Blackbum, Victoria 3130 895.2895 Hewlett-Packard Co. 301 E.Evelyu Mountain View, CA 94039 (415) 694-2000 IN CANADA Hewlett-Packard (Canada) Ltd.
Returning Instruments for Service If a module is being returned to Hewlett-Packard for servicing, fill in and attach a blue service tag. Service tags are supplied at the end of this manual. Please be as specific as possible about the nature of the problem. Include copies of error messages, data related to module performance, type of system, etc., along with the module being returned. Packaging The original shipping containers should be used.
a. Wrap the instrument in anti-static plastic to reduce the possibility of damage caused by ESD. b. For instruments that weigh less than 54kg (120-lb), use a double-walled, corrugated cardboard carton of 159-kg (350-lb) test strength. c. The carton must be large enough to allow three to four inches on all sides of the instrument for packing material and strong enough to accommodate the weight of the instrument. d.
Figure l-4.
01 Figue 1-5.
Installation Introduction This chapter contains information needed to install the HP 70903A into an HP 70000 Series mainframe and to configure and cable the HP 70903A for use in an HP 71100A or 71210A spectrum analyzer. These examples cover the HP 70903A cabling requirements for either a single or dual IF section configuration. Cabling for single- and dual-IF systems is shown in Figure 2-3 and Figure 2-4, respectively.
Addressing the Module The HP 70903A needs an appropriate HP-MSIB (Hewlett-Packard Modular System Interface Bus) address to allow communications with the system master. The HP 70903A module HP- ‘MSIB address is set using the module ’s ROW and COLUMN address switches. Determining the HP-MSIB Address Table 2-l The HP 70903A has a factory-preset HP-MSIB address of 2, 18 (row 2, column 18). shows the address map for an HP 71210A Microwave Spectrum Analyzer.
Setting the HP-MSIB Address Switches A module-address change requires the following steps: 1. Locate the address switches on the right side-panel of the module. See Figure 2-l for an example of the switches. 2. Set the address switches labeled “row ” to the binary value of the module ’s HP-MSIB row number. For example, if the row value is 2, set the switches to binary 010 as shown in Figure 2-l. 3. Set the five switches labeled “column ” to the binary value of the module ’s HP-MSIB column number.
Installing the Module in the Mainframe The HP 70903A must be installed in an HP 70000 Series mainframe. This is accomplished by following the procedure below. Figure 2-2 identifies the module and mainframe parts called out in the procedure. To avoid blowing the mainframe line fuse or internal module fuses, the mainframe power must be turned off before connecting or disconnecting modules. Caution # Module Installation 1. Turn the mainframe LINE switch off. 2 . Open the mainframe front-panel door. 3 .
Connecting the Rear-Panel Cables This section contains instructions for connecting the rear-panel cables to an HP 70903A in either an HP 71100A or an HP 71210A spectrum analyzer single-IF ’and dual-IF section system, respectively. The cables supplied with an HP 70903A module that was ordered separately allow it to be installed as shown in the examples. Longer cables must be ordered for other configurations.
Rear-Panel Cabling for HP 71210A Microwave Spectrum Analyzer Figure 2-4 depicts the rear-panel connections of an HP 71210A Microwave Spectmm Analyzer, which includes both an HP 70903A and an HP 70902A. Using Table 2-2 for identification and Figure 2-4 as an example, connect the IF rear-panel cables as described below. HP 71210A Dual-IF Section Cable Connections HP 70903A cable installation n Connect the RF section 21.4MHz OUTPUT and the HP 70903A IF Section 21.4MHz INPUT using the 7/8 span cable.
Specifications There are no specifications for the HP 70903A IF Sections. For system-level specifications and characteristics, refer to the installation and verification manual for the system master.
Verification Introduction This chapter normally contains module performance-verification tests, which evaluate the electrical performance of the module against its specifications. Since there are no module specifications for the BP 7090314 IF Section, no module performance-verification tests apply. Firmware The HP 70903A IF Section will work with any HP 70900A Local Oscillator (LO) firmware versions. However LO firmware dated 889314 and later provides a different menu display than older firmware.
System Diagnostics System Diagnostics is a downloadable (DLP) program that is downloaded into each standard HP 70090 Series system shipped to a customer. A DLP is a sequence of HP-IB codes which resides in the RAM of the LO and can be executed from within the analyzer at the press of a key. DLPs do not go away when power is turned off, so they can be put in an instrument and used repeatedly.
Spectrum Analyzer Operation Verification It is possible to verify the electrical performance of the entire spectmm analyzer, including log fidelity and resolution bandwidth of the HP 70993A IF Section, by using System Verification Programs. Refer to the installation and verification manual for the system master regarding system verification programs.
5 Troubleshooting Introduction This chapter provides information on the front-panel status indicator lights and the error messages produced by the HP 70903A IF Section. If problems are encountered, review Chapter 2 of this manual to ensure the module addressing switches are properly set, cabling is correct, and the module is securely seated in the mainframe.
STATUS ACT The ACT LED of a slave module is only operative when there is a display in the system or when the instrument is performing a self-test. Note The ACT LED of a slave module turns on when: n n n The module is performing a function directed by manual control of the display keyboard. The module is making a measurement and its master has keyboard control of the display. The instrument is in self-test mode.
Error Messages Spectrum-analyzer error-messages generated by an HP 70903A are listed in this section. The messages are grouped by functional category; each category has its own series of numbers. Refer to the installation and verification manual for the system master for a complete list of all system error-messages. Table 5-l.
Index A Accessories l-4 ACT LED 1-5, 5-l Addressing HP-MSIB 2-2 the module 2-2 C Cabling 2-5 dual IF Sections 2-6 single IF Section 2-5 D downloadable programs, 42 E ERROR LED 1-5, 5-2 Error messages 53 ESD heel-strap l-8 protection l-8 static mat$oor l-8 static mat;table l-8 wrist-strap l-8 F Fiiare 4 1 I Installation 2- 1, 2-4 M Manual updating l-4 Module self-test 5-2 P Padraging containers 1-12 damage l-4 material; modules l-15 materiak; mainframe 1-14 PC board handling l-9 PC board storage l-9 Prep
S Serial number label l-3 prefix l-4 Setting address switches 2-3 shipping containers 1-12 Status Indicator ACT, 41 ERR, 41 system diagnostics, 42 T test calibration, 43 operation verification, 43 system diaguoetics, 42 system performauce, 43 Test Self-Test, 41 Index-2