Specifications
4-8
Operating Features and Specifications
Clock and Clock (inverted) Outputs
Error Add
Add errors to the data using the key and its menu of softkeys. There are three
modes of operation:
Single: Adds single errors on demand.
Fixed: Fixederrorratiosof1errorin10
n
bits, n = 3, 4, 5, 6, 7, 8, 9.
External: Injects a single error in the transmitted test pattern on each rising edge at
the ERROR INJECT INPUT port.
Clock and Clock (inverted) Outputs
The following clock output features can be set using the key and its menu
of softkeys.
• Clock high level adjust.
• Clock amplitude adjust.
• External termination voltage 0/−2V or ac coupled.
• External attenuator set 0 to 40 dB (0V termination only).
• Independent control of high level and amplitude for CLOCK and CLOCK.
Specifications
Clock outputs - main
See figures 4-1 and 4-2 for amplitude and high-level specifications with 0V and −2V
termination.
Frequency range: 100 MHz to 12.5 GHz.
Interface: Complementary, dc coupled, 50Ω, reverse terminated.
Amplitude: 0.3 to 2 V pp in 10 mV steps.
Range: +1.5 to −3.0 V in 10 mV steps.
Connector: APC-3.5 mm male connector.
error add